Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27881
Brand Name:Straumann Pty Ltd - Screwdriver, surgical
Model:
Affected:Lots - EGT63
Software version:
Recalling Organisation:Straumann New Zealand Ltd
Contact Information:
Manufacturer:Institut Straumann AG
Issue:The manufacturer has received several complaints from customers that the SCS Screwdriver article 046.402 from lot FAP78 does not fit. The investigation of these complaints has confirmed the SCS tip of the screwdriver does not have the correct configuration / profile due to a manufacturing error therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/6/2021