Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27851 |
| Brand Name: | ACE Control Se |
| Model: | |
| Affected: | REF 1667001 (lots 80516, 80635, 90092, 90599, 00118, 00360, 00610, 00702) |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | ACE Control Set, failed to reach their stability claim after reconstitution of the quality control material. An upward shift has been observed during testing, up to a maximum of 25.5% of Level 1 and up to 12.5% of Level 2 with lot 90092. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/6/2021 |