Detail of Recall

Type of Product:Device
Medsafe Reference:27847
Brand Name: Patient monitor module, multifunction
Model:CARESCAPE PDM (New) - 2042084-001, CARESCAPE PDM (Goldseal) - 2094504-001
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems Information Technologies Inc
Issue:The manufacturer has identified that if the CARESCAPE PDM with Masimo SET technology has not been powered down completely in more than 2 years, it can display a frozen SpO2 saturation value that is inaccurate and no longer changes to reflect the patient's clinical condition. If this situation were to occur, it could result in missed SpO2 alarms and delayed diagnosis and treatment of hypoxia or hyperoxia.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/6/2021