Type of Product:Device
Medsafe Reference:27836
Brand Name:Avalon Fetal Monitors
Model:Philips Avalon Fetal Monitors FM20/FM30/FM40/FM50
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medizin Systeme Boblingen GmbH
Issue:The manufacturer has identified a software anomaly in the affected Fetal Monitors that can lead to any of the following reporting issues when generating a non-stress test (NST) report according to the Dawes & Redman algorithm:
1. Instead of the average baseline value, the maximum baseline value is printed on the NST report.
2. Delayed generation of a criteria met report.
3. A criteria not met report is generated instead of a criteria met report.
Reviewing the incorrect report can lead to misinterpreting the NST report. Patient harm is not expected to occur.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:4/6/2021