Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27835 |
| Brand Name: | CN - DBS StimLoc |
| Model: | StimLoc Burr Hole Cover - model number - 924256 and Spare Parts Kit for StimLo Burr Hole Cover - model number - 3550S-01 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | The manufacturer has identified that specific lots of the StimLoc clips may have excess nylon moulding flash. This issue can prevent the clip from snapping into the StimLoc base, which could allow for lead movement or lead damage. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 4/6/2021 |