Detail of Recall

Type of Product:Device
Medsafe Reference:27810
Brand Name:CRDM - Reveal LINQ - Insertable Cardiac Monitoring System
Model:
Affected:Product code - LNQ11
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:The manufacturer has identified that Reveal LINQ with TruRhythm ICMs that undergo a partial electrical reset appear to be programmed ON, but are no longer able to detect and report Brady and Pause events. A partial electrical reset is normal behaviour that can occur when the device detects a possible issue with the device software. However, an error in the partial electrical reset implementation is causing this unintended behaviour.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/5/2021