Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27810 |
| Brand Name: | CRDM - Reveal LINQ - Insertable Cardiac Monitoring System |
| Model: | |
| Affected: | Product code - LNQ11 |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | The manufacturer has identified that Reveal LINQ with TruRhythm ICMs that undergo a partial electrical reset appear to be programmed ON, but are no longer able to detect and report Brady and Pause events. A partial electrical reset is normal behaviour that can occur when the device detects a possible issue with the device software. However, an error in the partial electrical reset implementation is causing this unintended behaviour. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/5/2021 |