| Type of Product: | Device |
| Medsafe Reference: | 27795 |
| Brand Name: | Ysio Max |
| Model: | |
| Affected: | Catalogue/Lot number – 11107464 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has received a complaint about an image assigned to a different patient during image recovery process. In the unlikely event of an interrupted connection to a MAX detector during an initiated image recovery, the previously acquired image (possibly belonging to the previous patient) might be displayed to the user instead of the currently acquired image. If this error remains undetected by the user, it could potentially impact diagnosis of the patient examined when the connection was interrupted. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/6/2021 |