Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27794 |
| Brand Name: | X-wing |
| Model: | |
| Affected: | X-Wing Base (35130514, various batches - 181799, 182098, 182629, 191428, 191435, 201953 |
| Software version: | |
| Recalling Organisation: | Smith & Nephew Ltd 621 Rosebank Road Avondale AUCKLAND 1026 |
| Contact Information: | 0800 807 019 |
| Manufacturer: | Enztec Limited |
| Issue: | An inherent design flaw on the X-Wing Base drawing resulted in mis-placement of the "0" mark on a scale used for the preliminary measurement of a harvested graft. This scale is only used for initial measurement of the harvested graft and thus the error does not affect the final bundle length or the graft tensioning function of the X-Wing. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/5/2021 |