Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27771 |
| Brand Name: | ADVIA Chemistry XPT System |
| Model: | |
| Affected: | Catalogue/Lot number – 10723034 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | The manufacturer has confirmed that the system behavior in all commercially available software versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) for the ADVIA Chemistry XPT system may not be aligned with customer expectations for the current default mixer failure detection setting of 2: Warning+STOP (option 2 in System Parameters) for the Reagent 1 and Reagent 2 mixer assemblies. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/5/2021 |