Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27753 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | Multiple Models |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has identified an issue found with the voxelwise worst clinical goals in the Robust evaluation module of RayStation 8B, 9A, 9B, 10A and 10B. Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/5/2021 |