Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27673 |
| Brand Name: | Chargers 230v assorted |
| Model: | GS777 Wall Transformer |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Welch Allyn Inc |
| Brand Name: | Pro BP Handheld BP Device |
| Model: | Pro BP 3400 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Ebos Group Ltd 108 Wrights Road Addington Christchurch |
| Contact Information: | 0800 18 17 16 |
| Manufacturer: | Welch Allyn Inc |
| Issue: | The manufacturer has identified a risk of an internal electrical short and possible electrical shock occurring if cleaning fluid ingresses into the power supply, and the power supply is plugged into the electrical outlet before the cleaning fluid has completely dried. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/5/2021 |