Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27663
Brand Name:W/A ProBP SureBP
Model: ProBP SureBP
Affected:
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Welch Allyn Inc
Brand Name:Chargers 230v assorted
Model:Chargers 230v assorted
Affected:
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Welch Allyn Inc
Issue:There is a risk of an internal electrical short and possible electrical shock occurring if cleaning fluid ingresses into the power supply, and the power supply is plugged into the electrical outlet before the cleaning fluid has completely dried.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/5/2021