Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27662
Brand Name:W/A Spirometer diagnostic
Model:
Affected:
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Medikro Oy
Issue:On January 20, 2021 Microsoft Corporation released MS Windows 10 patch KB4577586 which removes Adobe Flash from the system. Once the patch is applied, it will uninstall the Adobe Flash Player and the spirometry module will not work. When the user selects the spirometry module, they will receive an error message that will return them to the main menu and a spirometry test cannot be performed.In both scenarios, users can still view existing spirometry exams stored in the database. Resting ECG is not impacted and will run without any issues.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/4/2021
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /