Detail of Recall

Type of Product:Device
Medsafe Reference:27613
Brand Name:Silk vista Intracranial self-expandable stent
Model:
Affected:All lots
Software version:
Recalling Organisation:Getz Healthcare Limited
Level 17
120 Albert Street
Auckland
Contact Information:0800 733 333
Manufacturer:Balt Extrusion SAS
Issue:The manufacturer received a complaint about an error in the MRI information regarding the Silk Vista stent. It was noticed in the Instruction For Use (IFU) that the SAR (Specific Absorption Rate) value for the Silk Vista is incorrect 0.5W/kg at 1.5 Tesla and 0.6W/kg at 3 Tesla. This level makes the use of MRI impossible, as in normal imaging conditions the SAR exceeds those values.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/4/2021