Detail of Recall

Type of Product:Device
Medsafe Reference:27579
Brand Name:IRON2
Model:on cobas c 311/501/502
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:Several customer complaints have been received regarding the increased recovery of controls and discrepant elevated results for the IRON2 on cobas c 311/501/502. Internal investigations confirmed the issue and revealed a systematic sample drift up to +4.7 µmol/L absolute for IRON2 over the entire measuring range. The bias increases with the number of tests performed from one cobas c pack without further calibration. The first measurements are not affected while the last sample can exhibit the maximal observed bias.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/4/2021