Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27576 |
| Brand Name: | Leckey Walking aids |
| Model: | |
| Affected: | Serial number - From LEC/220568 to LEC/272111 |
| Software version: | |
| Recalling Organisation: | Allied Medical Ltd 29 Triton Drive North Harbour Auckland |
| Contact Information: | 0800 31 61 81 |
| Manufacturer: | James Leckey Design Ltd |
| Issue: | Reports have indicated that the boss (B), has become loose from the tube (A), causing the castor to loosen from the rear leg of the chassis. The manufacturer investigations have confirmed that this is due to variability in the weld process used to attach the boss (B) to tube (A). In some instances, this has resulted in an insufficient weld between the leg and boss. |
| Recall action type: | Correction |
| Recall action: | |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/4/2021 |