Detail of Recall

Type of Product:Device
Medsafe Reference:27576
Brand Name:Leckey Walking aids
Model:
Affected:Serial number - From LEC/220568 to LEC/272111
Software version:
Recalling Organisation:Allied Medical Ltd
29 Triton Drive
North Harbour
Auckland
Contact Information:0800 31 61 81
Manufacturer:James Leckey Design Ltd
Issue:Reports have indicated that the boss (B), has become loose from the tube (A), causing the castor to loosen from the rear leg of the chassis. The manufacturer investigations have confirmed that this is due to variability in the weld process used to attach the boss (B) to tube (A). In some instances, this has resulted in an insufficient weld between the leg and boss.
Recall action type:Correction
Recall action:
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/4/2021