Detail of Recall

Type of Product:Device
Medsafe Reference:27575
Brand Name: CT Lightspeed Family
Model:Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT Apex Edition
Affected:
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:The manufacturer has become aware of a potential issue on Revolution CT, Revolution CT ES, Revolution Apex, Revolution CT Apex Edition systems when adding one or more intended scan groups. The additional group names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition. Consequently, each additional scan group with non-alphanumeric characters that the user attempts to add will be included in the acquisition but will not be visible to the user as being included in the scan. This could result in exposing the patient to unintended X-ray radiation.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/4/2021