| Type of Product: | Device |
| Medsafe Reference: | 27558 |
| Brand Name: | Artis Q, Q.zen, zeego, zee systems |
| Model: | |
| Affected: | Catalogue/Lot number – 10094137, 10502502 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that some Artis zee ceiling systems show an increased abrasion of the cabling at the cable outlet of the inner C-arm. This issue occurs sporadically and is not considered as a systematic issue. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/4/2021 |