Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27548
Brand Name:Med-con_masks[35177]
Model:All batches of Med-Con facemasks with a Med-Con logo on the corner of the mask
Affected:
Software version:
Recalling Organisation:Thermo Fisher Scientific New Zealand Ltd
244 Bush Road
Albany
AUCKLAND
Contact Information:0800 636 327
Manufacturer:Med-Con Pty Ltd.
Issue:The manufacturer has supplied five batches of Surgical Face Masks which, when tested by the TGA, demonstrated inconsistent testing results for fluid resistance at the location of the logo only. In some cases, these devices did not pass at level 1 at the location of the logo. Testing confirms the devices offer level 2 or 3 fluid resistance at the main body of the mask in line with the level of fluid resistance claimed.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/4/2021