Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27498
Brand Name:camera fundus 10551
Model: ZEISS CLARUS models 500 and 700 with software 1.1.1.
Affected:
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Meditec Inc
Brand Name:camera fundus 10551
Model:ZEISS CLARUS 500 and 700 devices with software version 1.1.1.
Affected:
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Meditec Inc
Issue:The manufacturer has become aware that if, at the same time of Auto Import completion, a user has selected a patient to be imaged and proceeds to acquire images for that patient, the selected patient may get de-selected without the user coming to know about the de-selection. This change of patient could lead to an acquired image to be placed under the wrong patient ID.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/5/2021