Detail of Recall

Type of Product:Device
Medsafe Reference:27487
Brand Name:SARS-CoV- & Influenza A/B kit
Model:
Affected:
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Roche Diagnostics GmbH
Issue:The manufacturer received complaints alleging invalid and/or false positive results with the cobas® SARS-CoV-2 & Influenza A/B Test for use on the cobas® Liat® System for one or more targets (SARS-CoV-2, Influenza A, Influenza B).
Update Version 2.
The manufacturer has announce the availability of new mandatory software update version 3.3.1 for the cobas® Liat® System and an updated assay script (v1.1.0) for cobas® SARS-CoV-2 & Influenza A/B Test.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/3/2021