Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27477 |
| Brand Name: | CVG - VenaSeal Adhesive Occlusion System |
| Model: | VenaSeal™ Closure System, SP-101 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | The manufacturer has determined that in some cases, the trigger of the VenaSeal dispensing gun, which advances the plunger rod forward resulting in forward movement of the syringe plunger and adhesive, does not advance and dispense adhesive as intended. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/3/2021 |