Detail of Recall

Type of Product:Device
Medsafe Reference:27467
Brand Name:ORTHO VISION Analyzer and ORTHO VISION Max Analyzer
Model:
Affected:Software versions - APSW 5.13.0 and below
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics
c/o Quigg Partners
Level 7, 36 Brandon Street
Wellington 6011
Contact Information:0800 402 899
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:The manufacturer has identified the potential for false negative results when using the ORTHO VISION® Analyzer and/ or ORTHO VISION® Max Analyzer with some specific User Defined Protocol configurations. This issue only affects customers processing UDP tests with non-standard volumes.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/3/2021