Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27454 |
| Brand Name: | CN - Percept PC BrainSense B35200 |
| Model: | Percept PC BrainSense B35200 - model number - B35200 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | This issue impacts Percept PC patients with cycling enabled. When using the replacement feature to transfer settings from an existing implanted device with cycling enabled to a newly implanted Percept device, cycling is not properly enabled, though it appears to be enabled in the clinician programmer screen and reports. As a result, the patient's stimulation is always on, and battery longevity may be significantly reduced. All Percept PC replacement patients previously implanted with Activa family of devices are at risk for premature battery depletion if they are configured with cycling using the A610 REPLACEMENT workflow. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 17/3/2021 |