Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27383
Brand Name:Diabetes - GLUCOSE SENSORS
Model:Enhanced Enlite™ Sensor, Guardian™ Sensor 3
Affected:
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic MiniMed
Issue:The manufacturer became aware about a medication named Hydroxyurea which could potentially impact sensor glucose reading. The presence of Hydroxyurea in patient's body could interfere with sensor chemistry, potentially resulting in inaccurately higher sensor glucose readings.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:5/3/2021