Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27368 |
| Brand Name: | Attune Rotating Platform Cemented Tibial Base |
| Model: | |
| Affected: | Lots - 9557376, 9557377,9558239 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (NZ) Limited 13a Gabador Place Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | DePuy Ireland |
| Brand Name: | Attune Fixed Bearing Cemented Tibial Base |
| Model: | |
| Affected: | Lots - 9557376, 9557377,9558239 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (NZ) Limited 13a Gabador Place Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | DePuy Ireland |
| Issue: | The manufacturer has received reports of packaging damage. Investigation of these complaints determined that product packaging may be compromised, which may present a potential for sterile barrier integrity failure, resulting in the possibility of non-sterile implant or residual packaging materials on the tibial implants. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/3/2021 |