Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27334 |
| Brand Name: | RaySearch_Radiation therapy treatment planning system |
| Model: | RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, 10A, 10B, including some of the service packs |
| Affected: | |
| Software version: | |
| Recalling Organisation: | CARSL Consulting 30 Fulford Road Havelock North 4130 Hastings |
| Contact Information: | 09 238 3447 |
| Manufacturer: | RaySearch Laboratories AB (publ) |
| Issue: | The manufacturer has found an issue with the Merge beams functionality in RayStation/RayPlan. For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labelled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/2/2021 |