Detail of Recall

Type of Product:Device
Medsafe Reference:27332
Brand Name: Hamilton Heater Shaker
Model:
Affected: Hamilton Robotics accessories (HHS – Hamilton Heater Shaker)
Software version:
Recalling Organisation:Bio-Strategy Ltd
22a William Pickering Drive
North Harbour
Albany
Auckland 0751
Contact Information:09 969 9150
Manufacturer:
Issue:The manufacturer has identified a small potential risk where the Hamilton Heater Shaker can start a fire if flammable liquids contact and enter the device through events such as missing labware during a pipetting step or overflow of fluid from labware or an on-deck fluid handling device.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/2/2021