Detail of Recall

Type of Product:Device
Medsafe Reference:27326
Brand Name:Alaris Plus Volumetric Pumps_BD Swiss
Model:BD Alaris GP pluswith plus software - BD Alaris GP plusGuardrails with plus software - BD Alaris neXus GP
Affected:
Software version:
Recalling Organisation:Becton Dickinson Ltd
Unit 4 / 110 Mays Road
Onehunga
Auckland 1061
Contact Information:0800 572 468
Manufacturer:BD Switzerland Sarl
Issue:It was reported to the manufacturer that pumps were generating false Flow Error alarms when a flow sensor was in use. The flow sensor is an ultrasonic device, which is attached to the drip chamber and detects if drops continue from the container into the drip chamber. If it detects no flow, a high priority alarm of 'No Flow' will be generated. If it detects too many drops, in relation to the programmed rate, a high priority alarm of 'Flow Error' will be generated
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/2/2021