Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27326 |
| Brand Name: | Alaris Plus Volumetric Pumps_BD Swiss |
| Model: | BD Alaris GP pluswith plus software - BD Alaris GP plusGuardrails with plus software - BD Alaris neXus GP |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Becton Dickinson Ltd Unit 4 / 110 Mays Road Onehunga Auckland 1061 |
| Contact Information: | 0800 572 468 |
| Manufacturer: | BD Switzerland Sarl |
| Issue: | It was reported to the manufacturer that pumps were generating false Flow Error alarms when a flow sensor was in use. The flow sensor is an ultrasonic device, which is attached to the drip chamber and detects if drops continue from the container into the drip chamber. If it detects no flow, a high priority alarm of 'No Flow' will be generated. If it detects too many drops, in relation to the programmed rate, a high priority alarm of 'Flow Error' will be generated |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/2/2021 |