Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27318 |
| Brand Name: | ENT - Nerve Integrity Monitors (NIM) |
| Model: | NIM4CM01 |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Xomed |
| Issue: | It has been observed that when the user is ready or needs to create a case report and selects the Interface Reports button, the system may show a black screen and reboot. After the system completes the reboot, the GUI is restored, and the user is back at the initial GUI screen for selecting a procedure type. Be aware that data from the first case performed after initial start-up of the system may not be available to print. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/2/2021 |