Detail of Recall

Type of Product:Device
Medsafe Reference:27316
Brand Name:Table _ Germany
Model:
Affected:Serial numbers - 1 - 3402, 1 - 1264, 1 - 401, 3 - 74, 2 - 1159, 2- 762
Software version:
Recalling Organisation:Getinge Australia Pty Ltd (NZ)
Unit D, Ground Floor, Building A, Millenium Centre
600 Great South Rd
Ellerslie
Auckland 1051
Contact Information:09 272 9039
Manufacturer:Maquet GmbH
Issue:Under certain conditions, the manufacturer has found that a problem may cause the operating table not to function as intended. The tables may develop a hydraulic leakage at the inclination function. This leakage leads to an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/2/2021