Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27316 |
| Brand Name: | Table _ Germany |
| Model: | |
| Affected: | Serial numbers - 1 - 3402, 1 - 1264, 1 - 401, 3 - 74, 2 - 1159, 2- 762 |
| Software version: | |
| Recalling Organisation: | Getinge Australia Pty Ltd (NZ) Unit D, Ground Floor, Building A, Millenium Centre 600 Great South Rd Ellerslie Auckland 1051 |
| Contact Information: | 09 272 9039 |
| Manufacturer: | Maquet GmbH |
| Issue: | Under certain conditions, the manufacturer has found that a problem may cause the operating table not to function as intended. The tables may develop a hydraulic leakage at the inclination function. This leakage leads to an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/2/2021 |