Type of Product:Device
Medsafe Reference:27312
Brand Name:Caesarea Medical Electronics Syringe Pump
Model:T34 3rd edition Syringe Pump
Affected:
Software version:
Recalling Organisation:Paragon Care Group New Zealand Limited
69 Elizabeth Knox Place
Glen Innes
AUCKLAND
Contact Information:0508 654 258
Manufacturer:Caesarea Medical Electronics Ltd (Israel)
Issue:The manufacturer, has become aware through customer feedback that if the T34 3rd Edition Ambulatory Syringe Pump is stored without the main 9V battery for several days or weeks, it may result in partial or full depletion of the 3V internal battery. The 3V battery powers the internal real-time clock. The 3V depletion may trigger two different issues: Timer's battery fail alarm and real-time clock time lag/delay
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/2/2021