Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27312 |
| Brand Name: | Caesarea Medical Electronics Syringe Pump |
| Model: | T34 3rd edition Syringe Pump |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Paragon Care Group New Zealand Limited 69 Elizabeth Knox Place Glen Innes AUCKLAND |
| Contact Information: | 0508 654 258 |
| Manufacturer: | Caesarea Medical Electronics Ltd (Israel) |
| Issue: | The manufacturer, has become aware through customer feedback that if the T34 3rd Edition Ambulatory Syringe Pump is stored without the main 9V battery for several days or weeks, it may result in partial or full depletion of the 3V internal battery. The 3V battery powers the internal real-time clock. The 3V depletion may trigger two different issues: Timer's battery fail alarm and real-time clock time lag/delay |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 19/2/2021 |