Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27304
Brand Name:Blood transfusion filter [35071] - PLT - IIb - HAE
Model:ATSBC1EE
Affected:
Software version:
Recalling Organisation:Haemonetics New Zealand Limited
No 120, 300 Richmond Road
Grey Lynn
Auckland 1021
Contact Information:
Manufacturer:Haemonetics Corporation
Brand Name:Blood transfusion filter [35071] - IIa - HAE
Model:RS1VAE
Affected:
Software version:
Recalling Organisation:Haemonetics New Zealand Limited
No 120, 300 Richmond Road
Grey Lynn
Auckland 1021
Contact Information:
Manufacturer:Haemonetics Corporation
Issue:The manufactuere has identified that the current convention for interpreting the expiry date for when only the year and month are printed on the label is to consider the product safe to use until the end of the printed month. During a software update in the production facility, the expiry date that was printed on the certificate was modified to show the YYYY-MM-DD format. This same modification was not carried over to the product labeling thus creating a discrepancy on the date format between the product labeling and the product certificate.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/2/2021