Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27295
Brand Name:CVG - APV - Valiant Navion Stent Graft
Model:
Affected:Multiple Product numbers
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Brand Name:CVG - APV - Valiant Navion Stent graft - US
Model:
Affected:Multiple Product numbers
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Inc
Issue:Update:
The manufacturer is currently working to further investigate and assess the cause of the events observed with the Valiant Navion Thoracic Stent Graft. To date, the root cause analysis suggests a loss of suture integrity, which could lead to separation of the longitudinal seam of the stent graft or stent ring detachment from the surface of the graft fabric.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/6/2022