Detail of Recall

Type of Product:Device
Medsafe Reference:27255
Brand Name:Biograph mCT Hoffman Estates
Model:Biograph Horizon systems running VJ21B
Affected:
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Medical Solutions USA Inc
Issue:The Manufacturer has new software update VJ21C that reduces the likelihood of a Versant Database user interface freeze which can result in CT rescan, specific to Biograph Horizon systems currently running VJ21B.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/2/2021