Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27243
Brand Name:Liko Overhead Lifting Devices
Model:MOD1322
Affected:
Software version:
Recalling Organisation:Universal Specialities Ltd trading as USL Medical
494 Rosebank Road
Avondale
AUCKLAND 1026
Contact Information:09 829 0595
Manufacturer:Liko AB
Issue:The manufacturer has receive complaints of incidents where the Multirall Q-link strap lock has not been fully engaged in the S65 rail carriage hook by user, leading to the Multirall lift becoming detached from the rail system, potentially resulting in the fall of both the lift motor and the patient.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/2/2021
Hide menus