Detail of Recall

Type of Product:Device
Medsafe Reference:27232
Brand Name:BioFire FilmArray
Model:
Affected:BD BACTEC™ Plus Aerobic medium in plastic vials - BD BACTEC™ PedsPlus medium in plastic vials
Software version:
Recalling Organisation:Medi'Ray New Zealand Ltd
53-55 Paul Matthews Road
Albany
Auckland 0632
Contact Information:0800 633 472
Manufacturer:Biomerieux SA
Issue:The manufacturer has identified an increased risk of false positive P. aeruginosa results when the BioFire BCID or BCID2 Panel is used with certain types of BD BACTEC™ blood culture vials with expiration dates of 31Jul2021 and 31Aug2021
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/2/2021