Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27193 |
| Brand Name: | Artis Q, Q.zen, zeego, zee systems |
| Model: | ARTIS icono and ARTIS pheno systems. |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | The manufacturer has identified that in some cases, when the DSA acquisition has been started while C-Arm or table is moving there might be a shift between the image acquired at start position and the image (i.e. DSA vessel map) taken at reference position. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/1/2021 |