Detail of Recall

Type of Product:Device
Medsafe Reference:27187
Brand Name:General Ultrasound Imaging System
Model:All EPIQ Systems (models EPIQ Elite, EPIQ 5G, EPIQ 5C, EPIQ 5W, EPIQ 7G, EPIQ 7C, EPIQ 7W, EPIQ CVx & EPIQ CVxi) shipped between 1/16/2019 and 10/22/2020.
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Ultrasound LLC
Issue:The manufacturer has identified that the EPIQ Ultrasound Systems could have 1-4 missing or loose screws on the base of the arm. There is a potential for the control panel arm to break and collapse if pressure is applied while adjusting or moving the control panel.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/1/2021