Detail of Recall

Type of Product:Device
Medsafe Reference:27163
Brand Name:V680 Ventilator
Model:
Affected:backup battery
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:The manufacturer has identified that the V680 ventilators may experience two potential issues with the battery.
1. The ventilator may exhibit a high priority Check Vent: Battery Failed alarm if the battery is allowed to over discharge.
2. The battery may fail to charge if the ventilator is stored for an extended period of time without being plugged into AC power.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/1/2021