Detail of Recall

Type of Product:Device
Medsafe Reference:27143
Brand Name:PrisMax
Model:
Affected:
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare SA (Switzerland)
Issue:The manufacturer identified that due to variability in the performance of the tubing in the ARPS (Automatic Repositioning System) Pump Assembly which may lead to alarm situations during a system self-test and during therapy. This tubing is normally replaced during annual preventative maintenance.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/1/2021