Detail of Recall

Type of Product:Device
Medsafe Reference:27109
Brand Name:ID-DiaCell
Model:
Affected:All lots currently in date
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Medical Diagnostics GmbH
Issue:The manufacturer has identifeid 2 issues when using ID-DiaCall:
1) NonSpecific reactions: - Primarily observed on QC samples but also with some patient samples and eluates. Doubtful reactions or in some cases false positive reactions may occur randomly between batches and within a same batch. This can appear with both manual and all automated methods.
2) Insufficient Homogenization of the RBCs within the vials when stored on-board either on the IH-500 or on the IH-1000 and when the remaining volume reaches 2ml or lower. This lack of homogenization can lead to a dispense of low-concentrated red blood cell reagent into the wells of the ID-cards.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/12/2020