Detail of Recall

Type of Product:Device
Medsafe Reference:27106
Brand Name:i-STAT CG8+ cartridge
Model:
Affected:i-STAT EG7+ cartridges - List number - 03P76-25
Software version:
Recalling Organisation:Abbott Diagnostics Division
Ground Floor, Building D
4 Pacific Rise
Mt Wellington
AUCKLAND 1640
Contact Information:0800 656 233
Manufacturer:Abbott Point of Care Canada Ltd
Issue:Through this internal testing process after storage at 30°C for 14 days, the manufacturer observed approximately 5.2% of cartridges reporting higher than expected results for ionized calcium. The manufacturer observed ionized calcium on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30°C) for periods of time in excess of:
3 days for lot numbers *20100 to *20339
7 days for lot numbers *20340 and above
Where * = lot letter A, K, L, M, N, W or Y.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/12/2020