Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27105 |
| Brand Name: | Alternating Systems - AtmosAir, Nimbus, Alpha, Active, Response, Logic and Trancell Ranges. Bari Breeze and Autologic 110 & 200; Therakair Visio; Auralis Mattresses & Seat Cushion |
| Model: | |
| Affected: | |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | The manufacturer has investigated customer complaints where mains cables have been damaged accidently or through miss-use (i.e. incorrectly stowed). Although the mains cables on the Alpha Active models listed have been tested to, and comply with IEC60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, and are fit for purpose, the investigation has identified ways of mitigating possible damage by increasing the cable wall thickness and incorporating an improved grommet. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/12/2020 |