Detail of Recall

Type of Product:Device
Medsafe Reference:27100
Brand Name:EVO IQ LVP
Model:
Affected:All versions below V01.04.00.01
Software version:
Recalling Organisation:Baxter Healthcare Ltd
33 Vestey Drive
Mt Wellington
AUCKLAND
Contact Information:0800 229 837
Manufacturer:Baxter Healthcare SA (Switzerland)
Issue:Update 13/05/24 - The manufacturer is providing an update on the activity to implement the hardware and software correction. The correction will be deployed via a product swap that will replace your current EVO version 1 pump with a newly manufactured pump with the sensor and software already installed (an EVO version 2 pump). The product code number will change to represent this change (from ELVP001ANZ to ELVP002ANZ).

Update 07/02/23: The manufacturer has been reviewing additional changes that could further prevent user misloading. A sensor has been designed for the EVO IQ LVP pumps to detect and mitigate improper administration set loading. Therefore, this corrective action will be extended to implement the sensor (a hardware fix) along with a corresponding software change (V01.07.00.01 or higher).
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/5/2024