Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27100 |
| Brand Name: | EVO IQ LVP |
| Model: | |
| Affected: | All versions below V01.04.00.01 |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare SA (Switzerland) |
| Issue: | Update 13/05/24 - The manufacturer is providing an update on the activity to implement the hardware and software correction. The correction will be deployed via a product swap that will replace your current EVO version 1 pump with a newly manufactured pump with the sensor and software already installed (an EVO version 2 pump). The product code number will change to represent this change (from ELVP001ANZ to ELVP002ANZ). Update 07/02/23: The manufacturer has been reviewing additional changes that could further prevent user misloading. A sensor has been designed for the EVO IQ LVP pumps to detect and mitigate improper administration set loading. Therefore, this corrective action will be extended to implement the sensor (a hardware fix) along with a corresponding software change (V01.07.00.01 or higher). |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 13/5/2024 |