Detail of Recall

Type of Product:Device
Medsafe Reference:27087
Brand Name:VITROS Chemistry Products TP and XT ALB-TP Slides
Model:
Affected:VITROS® Chemistry Products TP Slides - All current and future and past lots, VITROS® XT Chemistry Products ALB-TP Slides - all current, future and past lots
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics
c/o Quigg Partners
Level 7, 36 Brandon Street
Wellington 6011
Contact Information:0800 402 899
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:The manufacturer was made aware of new information regarding the potential for Eltrombopag to interfere with specific assays. During investigational testing by Ortho, VITROS TP and XT ALB-TP Slides were identified as sensitive to Eltrombopag interference.
Eltrombopag is a spectral interferent at the wave lengths that these VITROS assays are measured.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/12/2020