Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27081 |
| Brand Name: | Acetabular Components |
| Model: | |
| Affected: | PINN.SECTOR ACET.CUP 48MM, 50MM, 52MM, 54MM, 56MM, 60MM, - All lots numbers. PINN SECTOR W/GRIPTION 48MM, 50MM,52MM, 54MM, 58MM, 60MM - All lots numbers |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (NZ) Limited 13a Gabador Place Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Depuy Orthopaedics Inc |
| Issue: | There is a potential of an out of specification condition (oversized minor diameter) on certain PINNACLE® CUP devices. This defect is limited to lots that were manufactured on one specific machine tool in the manufacturing facility from July 2017 to November 2020. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/12/2020 |