Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:27080
Brand Name:Liquid Unassayed Multiqual
Model:Liquid Unassayed Multiqual, Liquid Assayed Multiqual, Liquichek Unassayed Chemistry Control
Affected:
Software version:
Recalling Organisation:Bio Assist Industries Pty Ltd
20 Mount View Close
Razorback
New South Wales 2571
Australia
Contact Information:
Manufacturer:Bio-Rad New Zealand
Issue:The manufacturer has receive customer report regarding lower calcium recovery observed in some vials, when the product is initially thawed, followed with recovery increasing to the expected values within several hours after thaw.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:21/12/2020
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /