Detail of Recall

Type of Product:Device
Medsafe Reference:27056
Brand Name:EliA dsDNA Well
Model:
Affected:Kit lot number 0142
Software version:
Recalling Organisation:Abacus dx New Zealand
Unit B, 30 Vestey Drive
Mt Wellington
Auckland 1060
Contact Information:0800222170
Manufacturer:Phadia AB
Issue:The manufacturer has identified that the EliA dsDNA Well lot BFA3Y/ 0142 gives higher results, and in some cases the results may be up to 3 times higher than for other EliA dsDNA Well lots. Results >15 IU/ml to 45 IU/ ml may potentially be false positive. The investigation has concluded that there are no deviations for the EliA ANA Positive Control or other EliA dsDNA Well lots.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/12/2020