Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 27054 |
| Brand Name: | Artis Q, Q.zen, zeego, zee systems |
| Model: | |
| Affected: | Artis zee/Q/Q.zen systems with software version VD11E |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare GmbH |
| Issue: | With the currently installed software there is a possibility of false error messages of motion control issues and jerky movements of the table hydraulic. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/12/2020 |